Abstract
Acetabulum cups are the counterparts of endoprostheses, which are mounted in the pelvis. There are different designs of such cups. They differ in their form and in the way of mounting. This paper deals with the threaded acetabulum cups with conical shape mounted in the pelvis. In general, one has to distinguish between stresses caused by the mounting process and lifetime stresses of the mounted cup. Stresses remaining of the mounting must ensure the so-called primary stability; however, they will disappear to some degree due to rheological properties of bone
The question arises whether the mounting stresses are responsible for local plastic deformations and the corresponding residual stresses or not. In this paper, mostly loading stresses and their dependence on the contact properties between pelvis and cup or between cup and polyurethane inlay are considered
چکیده
کاپ های استابولوم نقطه مقابل پروتز داخلی که در لگن نصب می شوند، می باشند. طراحی های متفاوتی از چنین کاپ هایی وجود دارد. آنها در شکلشان و طریقه نصب تفاوت دارند. این مقاله در رابطه با کاپ استابولوم پیچ و مهره ای، نصب شده در لگن با شکل مخروطی می باشد. معمولا باید تمایزی بین تنشهای ایجاد شده توسط فرآیند نصب و تنشهای طول عمر کاپ نصب شده باشد. تنشهای باقیمانده از نصب باید اصطلاحا پایداری اولیه را تضمین کنند؛ هر چند آنها در بعضی مراتب به دلیل خصوصیات رئولوژیکی استخوان ناپدید می شوند.